subperiosteal implants
Additively manufactured titanium scaffolds and osteointegration – meta-analyses and moderator-analyses of in vivo biomechanical testing
- 10 June 2021
- Posted by: Subperiosteal Institute
- Category: Study on Osseointegration
Simon Cleemput, Stijn E F Huys, Robbert Cleymaet, Wilfried Cools, Maurice Y Mommaerts
Full text link: https://pubmed.ncbi.nlm.nih.gov/34112248/
Additively Manufactured Titanium Scaffolds and Osteointegration: Meta-Analyses and Moderator Analyses of In Vivo Biomechanical Testing
1. Scientific Reference
- Study Title: Additively Manufactured Titanium Scaffolds and Osteointegration: Meta-Analyses and Moderator Analyses of In Vivo Biomechanical Testing
- Authors: Simon Cleemput, Stijn E. F. Huys, Robbert Cleymaet, Wilfried Cools, Maurice Y. Mommaerts
- Journal: Biomaterials Research
- Year of Publication: 2021
- DOI: 10.1186/s40824-021-00216-8
2. Scientific Background
Three-dimensional printing has expanded the possibilities for designing titanium implants intended for bone reconstruction and implant-supported rehabilitation. Unlike conventional manufacturing techniques, additive manufacturing allows the production of highly controlled porous structures with interconnected internal architectures that may enhance biological integration with surrounding bone.
Within implantology and reconstructive surgery, porous titanium scaffolds have attracted considerable attention because their architecture can potentially improve osteointegration while simultaneously influencing the mechanical behavior of the implant. Numerous experimental studies have evaluated variables such as pore size, overall porosity, lattice geometry, and post-processing surface modifications. However, the growing body of literature has generated a wide range of reported outcomes, making it difficult to identify which design characteristics consistently contribute to improved bone-implant integration.
This study addresses that challenge by systematically examining available animal data and quantitatively comparing the biomechanical performance of additively manufactured porous titanium implants.
3. Study Objective
The primary objective of the study was to evaluate how different additively manufactured porous titanium scaffold designs and post-printing treatments influence osteointegration outcomes measured through biomechanical implant-removal testing in animal models.
The authors also sought to determine whether specific scaffold characteristics—including pore size, porosity, lattice design, and post-processing treatments—could statistically explain variations in the strength of the bone-titanium interface observed across studies.
4. Methodology
This investigation was designed as a systematic review combined with two meta-analyses and multivariable meta-regression analyses. Literature searches were conducted in PubMed and Embase following CAMARADES and PRISMA recommendations.
Only in vivo animal studies assessing osteointegration through quantitative mechanical implant-removal tests were included. Eligible studies used push-out, pull-out, or torque-out testing methods to evaluate the strength of the bone-implant interface.
Two separate meta-analytical datasets were created:
- Porous 3D-printed titanium versus dense titanium.
- Post-treated porous titanium versus untreated porous titanium.
After screening and eligibility assessment, 10 studies were included in the first meta-analysis and 14 studies in the second. The included investigations involved several animal models, including rats, rabbits, dogs, and sheep.
5. Main Results
The analyses demonstrated that porous additively manufactured titanium implants generally exhibited superior biomechanical fixation compared with dense titanium controls. Similarly, post-printing modifications applied to porous titanium scaffolds were associated with improved mechanical removal strength when compared with untreated porous structures.
Among all variables investigated, two factors emerged as statistically significant moderators:
- Animal model used for implantation.
- Type of post-printing treatment applied to the scaffold.
A subgroup analysis focused on rabbit models, which represented the largest proportion of available data, identified a significant interaction between pore size and porosity.
The study also reported that calcium-containing coatings did not improve implant removal strength, whereas several other post-processing approaches appeared to provide beneficial effects.
Variables such as implantation time and implant location were not found to have statistically significant effects within the analytical models.
6. Clinical Analysis
The value of this study lies in its attempt to move beyond individual experimental reports and identify broader patterns across the rapidly expanding field of additively manufactured porous titanium implants.
Rather than defining a single optimal scaffold design, the findings suggest that osteointegration is influenced by a complex interaction of architectural and biological variables. The observation that animal species significantly affected outcomes highlights the importance of cautious interpretation when translating preclinical findings into human clinical applications.
The interaction identified between pore size and porosity further indicates that scaffold performance cannot be explained by one design parameter alone. This is particularly relevant for implantology, oral surgery, and maxillofacial reconstruction, where customized titanium implants are increasingly produced through digital workflows and CAD/CAM technologies.
Another important contribution of the study is its emphasis on methodological variability. Differences in scaffold design, implantation sites, testing protocols, and reporting standards continue to limit direct comparison between studies. Consequently, while porous titanium structures appear promising for enhancing biomechanical fixation, the current evidence remains insufficient to establish definitive design recommendations.
The findings therefore support continued research efforts aimed at standardizing scaffold characterization and biomechanical evaluation protocols.
7. Clinical Applications
Although the included evidence is preclinical, the findings may be relevant to several areas of implant-based rehabilitation and reconstructive surgery.
Potential applications include:
- Development of patient-specific titanium implants produced through additive manufacturing.
- Bone reconstruction procedures requiring enhanced bone-implant integration.
- Advanced implantology workflows involving digitally designed implant structures.
- Maxillofacial and craniofacial reconstruction using porous titanium frameworks.
- Future optimization of implant-supported prosthetic rehabilitation systems.
Because all included studies were conducted in animal models, direct clinical extrapolation should be approached cautiously.
8. Level of Evidence, Limitations and Transparency
This publication represents a systematic review with meta-analyses and moderator analyses, providing a relatively high level of evidence within the context of preclinical research.
However, several limitations should be considered. The included studies demonstrated substantial heterogeneity regarding animal species, scaffold characteristics, implantation protocols, and biomechanical testing methods.
The authors also reported funnel plot asymmetry, suggesting the possibility of publication bias.
No specific conflicts of interest are reported within the information provided. Likewise, detailed disclosures regarding potential industry influence are not available in the extracted material reviewed here.
As with all animal-based evidence, caution is required when interpreting the clinical significance of the findings.
9. Key Study Points
- Study type: Systematic review with meta-analysis and meta-regression.
- Level of evidence: High-level preclinical evidence.
- Number of studies analyzed: 10 studies in Meta-analysis 1 and 14 studies in Meta-analysis 2.
- Number of patients/implants: Not specified in the provided data.
- Follow-up period: Variable across studies, ranging from approximately 2 weeks to 6 months.
- Primary outcome: Mechanical strength of the bone-titanium interface measured through implant-removal testing.
- Main finding: Porous additively manufactured titanium generally demonstrated stronger biomechanical fixation than dense titanium.
- Scientific conclusion: Animal model and post-processing treatment significantly influenced observed outcomes.
- Potential limitations: Study heterogeneity and evidence of possible publication bias.
10. Scientific Impact
This study provides one of the first quantitative syntheses specifically focused on the biomechanical osteointegration performance of additively manufactured porous titanium scaffolds.
Its contribution extends beyond confirming the advantages of porous titanium structures. By identifying moderators associated with outcome variability, the research helps explain why previous experimental studies have often produced differing conclusions. The work also highlights the importance of considering scaffold architecture as a multidimensional concept rather than focusing on isolated parameters such as pore size alone.
From a research perspective, the findings support the need for improved reporting standards and greater consistency in experimental design. Such standardization could facilitate more meaningful comparisons between studies and accelerate the development of optimized scaffold designs.
As additive manufacturing continues to play a growing role in implantology, oral surgery, and personalized reconstruction, this study offers an important framework for future preclinical investigations.
11. Editorial Conclusion
This systematic review and meta-analysis provide a comprehensive assessment of current preclinical evidence regarding additively manufactured porous titanium scaffolds. The findings support the concept that porous titanium architectures can improve biomechanical bone fixation compared with dense titanium, while also demonstrating the influence of post-processing treatments and experimental models on reported outcomes. Although the results are encouraging, the substantial heterogeneity of available studies and the exclusively animal-based evidence base warrant careful interpretation before translating these findings into routine clinical practice.