subperiosteal implants
Comparison of Screws with Different Diameters in Subperiosteal Implant Application with Finite Element Analysis
- 7 October 2024
- Posted by: anjaform
- Category: Finite Element Studies
Abdulsamet Kundakcioglu, Betul Gedik
Full text link: https://pubmed.ncbi.nlm.nih.gov/39439467/
Comparison of Screws with Different Diameters in Subperiosteal Implant Application with Finite Element Analysis
1. Scientific Reference
- Study Title: Comparison of Screws with Different Diameters in Subperiosteal Implant Application with Finite Element Analysis
- Authors: Abdulsamet Kundakcioglu, Betul Gedik
- Journal: International Journal of Medical Sciences
- Year of Publication: 2024
- DOI: 10.7150/ijms.93225
2. Scientific Background
The renewed interest in subperiosteal implants reflects the growing demand for treatment options in patients with advanced jawbone atrophy, where conventional endosseous implant placement may be challenging or require extensive bone augmentation procedures. Advances in digital imaging, CAD/CAM technology, and patient-specific manufacturing have enabled the development of customized subperiosteal implant designs that better conform to individual anatomical conditions.
Within this context, the biomechanical behavior of the fixation system plays a critical role in the long-term stability of the implant. The diameter of fixation screws may influence stress distribution within the supporting bone and affect the overall mechanical response of the implant assembly. This study investigates these biomechanical aspects using finite element analysis (FEA), a computational modeling technique widely employed to evaluate implant performance under simulated functional loading before clinical implementation.
3. Study Objective
The primary objective of this study was to evaluate how different fixation screw diameters influence stress distribution in the supporting bone and the mechanical behavior of customized subperiosteal implants. Using finite element analysis, the authors compared 1.5 mm and 2 mm screws to determine which diameter provided more favorable biomechanical support under simulated masticatory loading.
4. Methodology
This investigation was designed as a preclinical computational study based on finite element analysis (FEA).
Computed tomography data were converted into STL files to generate two three-dimensional skull models. Customized subperiosteal implants were designed and positioned on these models. The researchers then simulated the placement of fixation screws measuring 1.5 mm and 2 mm in diameter. A chewing force of 250 N was applied to evaluate bone stress, implant stress, and implant displacement under functional loading conditions.
The study did not involve human participants or clinical implant placement. Information regarding patient numbers, follow-up duration, or clinical outcomes was not applicable and was not reported in the available abstract.
5. Main Findings
The finite element simulations demonstrated measurable biomechanical differences between the two screw diameters evaluated.
The 2 mm screws generated lower residual stresses within both the supporting bone and the subperiosteal implant compared with the 1.5 mm screws, suggesting a more favorable mechanical load distribution under the simulated conditions.
Conversely, the 1.5 mm screws exhibited less implant displacement during loading, indicating improved positional stability within the computational model.
Considering all evaluated parameters, the authors concluded that the 2 mm screw diameter appeared to provide the overall biomechanical advantage. However, they emphasized that these findings represent an initial step in investigating this treatment concept and should not be interpreted as clinical evidence supporting one fixation strategy over another.
6. Clinical Analysis
Rather than providing direct clinical evidence, this study offers insight into the mechanical behavior of customized subperiosteal implant fixation through computational modeling. The findings suggest that screw diameter may substantially influence how functional loads are distributed between the implant framework and the supporting bone.
From a biomechanical perspective, reducing stress concentrations is generally considered desirable because excessive localized stress may compromise the mechanical environment surrounding implant-supported reconstructions. The improved stress distribution observed with 2 mm screws may therefore represent a potentially favorable design characteristic. At the same time, the lower implant displacement associated with 1.5 mm screws illustrates that implant stability cannot be assessed using a single mechanical parameter alone.
For clinicians involved in complex implant rehabilitation, particularly in patients with severe alveolar bone atrophy where customized subperiosteal implants may be considered, this study highlights the value of digital engineering tools during implant design and optimization.
Nevertheless, these results should be interpreted cautiously. The simulations were performed under controlled computational conditions using a single loading scenario and cannot reproduce the complexity of biological healing, bone remodeling, patient variability, or long-term functional loading encountered in clinical practice.
7. Clinical Applications
The findings may contribute to the optimization of patient-specific subperiosteal implant designs developed through digital workflows integrating CT imaging, CAD/CAM technology, and customized manufacturing.
This research may also be relevant for clinicians and researchers investigating implant rehabilitation strategies for patients with advanced maxillary or mandibular bone atrophy, particularly where conventional implant placement is difficult.
However, the available evidence does not support modifying clinical protocols or recommending a specific fixation screw diameter solely on the basis of this computational study. Clinical validation remains necessary before these biomechanical observations can influence routine treatment planning.
8. Level of Evidence, Limitations and Transparency
This publication represents a preclinical finite element analysis, which provides valuable biomechanical information but constitutes a relatively low level of clinical evidence.
The conclusions are based entirely on computational simulations rather than clinical outcomes. Consequently, the study cannot evaluate biological responses, implant survival, complication rates, or long-term treatment success.
The authors acknowledge that their investigation represents an early step in exploring the potential of this treatment approach.
Beyond the inherent limitations associated with computational modeling, additional methodological limitations are not detailed in the available abstract.
No conflicts of interest or industry affiliations are reported within the information provided.
9. Key Study Points
- Study design: Preclinical finite element analysis.
- Level of evidence: Computational biomechanical study.
- Population/models: Two three-dimensional skull models.
- Number of patients or implants: Not specified in the available information.
- Follow-up: Not applicable.
- Primary outcome: Bone stress, implant stress, and implant displacement according to screw diameter.
- Main finding: Two-millimeter screws reduced residual stress within both bone and implant compared with 1.5 mm screws, while the smaller screws demonstrated less implant displacement.
- Scientific conclusion: Within this computational model, 2 mm fixation screws showed more favorable overall biomechanical performance.
- Limitations: Simulation-based study without clinical validation; additional methodological limitations are not detailed in the available abstract.
10. Scientific Impact
This study contributes to the growing body of research focused on the biomechanical optimization of customized subperiosteal implants. Rather than evaluating clinical outcomes, it provides computational evidence that fixation screw diameter can influence stress distribution and implant behavior under simulated functional loading.
The work supports the integration of finite element analysis into the digital design process for complex implant-supported rehabilitations, where engineering optimization may improve implant performance before clinical application.
Importantly, the study does not establish clinical superiority of one fixation strategy over another. Instead, it provides preliminary biomechanical data that may guide future experimental investigations and prospective clinical studies evaluating the long-term performance of customized subperiosteal implant systems.
11. Editorial Conclusion
This preclinical investigation demonstrates the usefulness of finite element analysis for assessing the mechanical performance of customized subperiosteal implant fixation systems. The computational models suggest that 2 mm fixation screws may offer a more favorable stress distribution than 1.5 mm screws, although smaller screws exhibited reduced implant displacement. Given the purely computational nature of the study, these findings should be regarded as preliminary biomechanical evidence requiring confirmation through experimental and clinical research before influencing routine implant treatment protocols.